MTS 793 Controller Hardware FlexTest GT Controller Connections
141
Single 407 Controller
Interlock Connections
The following figure shows how to cable interlocks for a single station
Model 407 Controller to the Model 493.10 Chassis (via its Model 493.74
HSM transition module). Use the specified interlock cable
(PN 056-455-0xx) to make these connections.
How to Jumper the
Hydraulic Terminal on a
Single Station 407
Controller
The following figure shows how to cable the hydraulic terminal on a
single station Model 407 Controller.
Run/Stop,
ActiveHigh
Hydraulic
Interlock,
ActiveLow
To
407
Controller
To
493.74
HSM
Transition
Interlock Cable PN 056-455-0XX
Digital
I/O Connector
Jumper
Jumper
Jumper
Hyd
In
Hyd
Out
Intlk
In
Intlk
Out
407 Controller
No Connection
Jumper Plug PN
049-635-901
Jumper Plug PN
049-635-901
No Connection