MTS 793 Controller Hardware
219
Adding a temperature
control channel to the
station configuration
file
After updating the hardware interface file, you must add a temperature
control channel to the Controller’s station configuration file.
To do this, use the Station Builder application to open the station
configuration file, then add temperature controller input (Temp Control
1-Input) and output (Temp Control 2-Output) resources to a Program
w/Feedback type channel. This creates a temperature control channel.
For detailed instructions, refer to “How to Program a Eurotherm
Temperature Controller” in the MTS 793 Control Software manual for
more information.