7-8 Remote control
7.6.2 SETTING
Commands
Behaviour
ASBK
This command aborts a sweep currently in progress previously started by command
SSFD. The reply is either:
• SBK=OK which means the sweep has been stopped and all setting
commands are available again.
• SBK=SERR which informs that the sweep was not in need to be stopped
ASPA
This command suspends a sweep currently in progress previously started by
command SSFD. There is no reply unless a sweep is not in progress in which case
the reply is:
• SPA=SERR which informs that the sweep was not in need to be paused
ASRE This command resumes a sweep previously paused by command ASPA. There is
no reply unless a sweep is not in progress in which case the reply is:
• SRE=SERR which informs that the sweep was not in need to be resumed
S3PRC This setting command #S3PRC * sets the PMM 7010 to conducted mode (BAND B).
The reply is:
• 3PR=OK (The PMM 7010 has been set for conducted mode )
• 3PR =BERR (The command has been ignored for the following:
• The instrument is not idling and therefore cannot switch to a different
mode. In such an event, stop the current process (Manual, Analyzer or
Sweep) and put it back to Idle.
• 3PR =SERR (The command contains a syntax error )
S3PRR This setting command #S3PRR * sets the PMM 7010 to radiated mode (BANDS C,
D and E). The reply is:
• 3PR=OK (The PMM 7010 has been set for radiated mode )
• 3PR =BERR (The command has been ignored for the following:
• The instrument is not idling and therefore cannot switch to a different
mode. In such an event, stop the current process (Manual, Analyzer or
Sweep) and put it back to Idle.
• 3PR =SERR (The command contains a syntax error )
SAAT a
This setting command sets Attenuator, for Analyzer Mode, to the value indicated by
(a) which can be 0 to 35 (55). Automatic Attenuator feature is disabled. If (a) is a
negative figure then the Automatic Attenuator feature is turned on. The reply is
AAT=OK which acknowledges the command has been granted or AAT =SERR if
the command has been ignored.
Example: # AAT 15* (set attenuator to 15dB and disable Automatic Attenuator
feature)
Example: # AAT -1* (Enable Automatic Attenuator feature)