4 About this device
RadMan 2LT/2XT 13
4.3 The operating modes
In addition to the standard Monitor operating mode of the RadMan 2LT,
RadMan 2XT offers the RF Detection operating mode.
Monitor RF Detection
(2XT only)
Application • Warning: Electromagnetic
radiation
• Detection of high-
frequency continuous
signals or pulsed
signals
• Leakage search on
waveguides
• Check to determine
whether antennas are
actually switched off
Integration time • 2LT: 1 s
• 2XT: 1 s of 30 ms
•30ms
Field exposure
display
• 5 – 200%, individual red
LED indicates
measurement value for E-
or H-field
(whichever is higher)
• 5 – 200%, red LEDs
as bar display
• Acoustic signal above
5%, sound becomes
higher with increasing
value
Alarm
threshold 1
50% exposure
and higher
1)
1) The alarm signals switch off once the limit values are no longer exceeded.
• Alarm LED flashes slowly
• Alarm signal sounds
slowly
• Slow vibration
• Alarm LED flashes
slowly
• Slow vibration
Alarm
threshold 2
100% exposure
and higher
1)
• Alarm LED flashes rapidly
• Alarm signal sounds
rapidly
• Fast vibration
• Alarm LED flashes
rapidly
• Fast vibration
Data logger • Activated • Deactivated