5 Operation
RadMan 2LT/2XT 23
Charge state < 5% remaining capacity
• flashes red.
The rechargeable batteries should be charged within the next 10 hours
(see Charging batteries on page 20).
When charging
During charging:
• flashes green.
• While switched off, the exposure LEDs indicate the already charged
capacity as a bar.
After completing the charging process:
• illuminates green.
• The exposure LEDs illuminate as a bar up to 100%.
5.5 Switching the integration time
(RadMan 2XT only)
In the Monitor operating mode, the integration time of the measurement can
be set to 1 s or 30 ms. For the reliable detection of brief or pulsed fields, the
short integration time of 30 ms is recommended.
Setting the integration time:
The device is switched on and in the Monitor operating mode (default).
After switching on, the integration time is set to 1 s.
Press 3x.
The integration time changes to 30 ms.
Three short beeps sound and the Pulse LED illuminates continuously.
Press 3x.
The integration time changes back to 1 s.
One long beep sounds.