5 Operation
RadMan 2LT/2XT 19
5.2 Power supply
As standard, power is supplied by two NiMH batteries of type AA (Mignon).
5.2.1 Operation with batteries
The device is delivered with partially charged batteries. The batteries must be
fully charged prior to the first use, however. Charging can be performed at the
USB-C connection via a suitable USB port (e.g., on a PC) or with a standard
USB charger (min. 500 mA charging current). The fully charged batteries can
supply power for an operating duration of approx. 800 hours (without alarm).
NOTE: Already a short battery charging time of approx. 5 minutes allows
the device to be used for up to 8 hours.
5.2.2 Operation with non-rechargeable batteries
NOTE: Operation with non-rechargeable batteries is not recommended,
but is, in principle, possible.
Regardless of the use of non-rechargeable batteries, charging is started
when a USB connection is made or when the charger is connected.
Therefore, be certain to observe the following warning instruction.
CAUTION
Use of non-rechargeable batteries
When using non-rechargeable batteries, establishing a USB
connector or connecting the charger results in an impermissible
charging process. This may cause the batteries to explode and lead
to severe injuries.
If non-rechargeable batteries are used, do not establish a USB
connection and do not connect the charger.
Only use non-rechargeable batteries if absolutely necessary and
replace with rechargeable batteries as quickly as possible.