11 Technical data
38 RadMan 2LT/2XT
Frequency response/isotropy
NOTE: Values in the positive range of the frequency response mean an
early warning.
Field RadMan 2LT RadMan 2XT
ICNIRP 1998, General Public
E ±3.5 dB (50 MHz to 8 GHz) Not available
ICNIRP 1998, Occupational (corresponds to Directive 2013/35/EU)
E ±3.5 dB (50 MHz to 8 GHz) • ±3 dB (900 kHz to 10 GHz)
• +6/-3 dB (> 10 to 20 GHz)
• +10/-3 dB (> 20 to 60 GHz)
H ±3 dB (50 MHz to 1 GHz) ±3 dB (27 MHz to 1 GHz)
FCC 96-326, Occupational
E ±3.5 dB (50 MHz to 8 GHz) • ±3 dB (3 MHz to 10 GHz)
• +6/-3 dB (> 10 to 20 GHz)
• +10/-3 dB (> 20 to 60 GHz)
H ±3 dB (50 MHz to 1 GHz) ±3 dB (3 MHz to 1 GHz)
Safety Code 6 (2015), Controlled
E • +4/-3 dB (50 MHz to 3 GHz)
• +6/-3 dB (3 GHz to 8 GHz)
• +4/-3 dB (10 MHz to 10 GHz)
• +6/-3 dB (> 10 to 20 GHz)
• +11/-3 dB (> 20 to 60 GHz)
H +4/-3 dB (50 MHz to 1 GHz) +4/-3 dB (27 MHz to 1 GHz)
Isotropic response
E±1dB (< 2.7GHz)
H ±1 dB (< 500 MHz)