NELLCOR
Sensors
CLEANING
AND
REUSE
Do
not
immerse
any
OXISENSOR, DURASENSOR,
OXIBAND,
or
RS-10 oxygen
transducer,
or
any
Nellcor
adhesive
in
water
or
cleaning
solution.
Clean
DURASENSOR,
OXIBAND,
and
RS-10 oxygen transducers by wiping
with
a
cleaning
agent
such
as
isopropyl alcohol. Do
not
sterilize by
irradiation,
steam,
or
ethylene
oxide.
Use
a
new
OXIBAND adhesive
wrap
for
each
patient.
Do
not
resterilize
OXISENSOR
oxygen
transducers.
PERFORMANCE
CONSIDERATIONS
If
there
is excessive
ambient
light, cover
the
sensor
site
with
opaque
material.
Failure
to do so
may
result
in
inaccurate
measurements.
light
sources
that
can
affect performance
include surgical lights (especially those
with
a xenon
light
source), bilirubin lamps, fluorescent lights,
infrared
heating
lamps,
and
direct
sunlight.
If
poor perfusion affects performance, consider
using
the
OXISENSOR
R-15
sensor;
it
obtains
measurements
from
the
nasal
septal
anterior
ethmoid artery,
an
artery
supplied by
the
internal
carotid.
This
sensor
may
obtain
measurements
when
peripheral perfusion is relatively poor.
For
low peripheral
perfiision, consider
using
the
NELLCORRS-10 sensor, which is ^
applied to
the
forehead or temple, sites
that
may
be
spared
during peripheral vasoconstriction. w
If
patient
movement
affects performance,
determine
whether
the sensoris properly and securely apphed,
move
the sensorto ^
a
less
active
site,
or
use
a
new
adhesive
sensor.