38 • Computer Control NEW FOCUS, Inc.
Set Standard Event Status Enable Register
Syntax *ESE <
value>
Vortex Actions Sets the specified bits of the Event Status Enable
Register.
When a corresponding bit in the Standard Event
Status Register becomes true, enables the Event
Summary Bit (ESB) in the Status Byte.
Argument An integer in the range 0 to 255, corresponding to
the combined weight of the bits you want to set.
Example Enable the ESB if the system is restarted:
*ESE 128
Standard Event Status Enable Register
Bit # Weight Description
7 128 Enable ESB in case of power on or
restart.
6 64 not used
5 32 IEEE: enable ESB in case of a com-
mand parse error.
RS-232: not used.
4 16 IEEE: enable ESB in case of system
not available or “invalid data
error.
RS-232: not used.
3 8 not used
2 4 Enable ESB in case of query error.
1 2 not used
0 1 not used