EasyManua.ls Logo

New Focus Vortex 6000 Series - Sense Laser Power; Set Piezo Voltage; Query Piezo Voltage Setpoint; Sense Piezo Voltage

New Focus Vortex 6000 Series
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6000 Vortex Series Computer Control • 47
Wavelength (Piezo Voltage) Commands
Sense Laser Power
Syntax :SENSe:POWer
Vortex Actions Returns the laser-power reading.
Response In constant-power mode:
“xx.x%”
range: 0.0 to 100.0
In constant-current mode:
“xx.xmW”
range: 0.0 to Max Power (laser specific)
Set Piezo Voltage
Syntax :SOURce:VOLTage:PIEZo <
voltage>
Vortex Actions Sets the piezo-voltage setpoint.
Argument x.x
range: 0.0 to 117.5, units are volts
Example Set the piezo voltage to 88.6 volts:
:SOUR:VOLT:PIEZ 88.6
Query Piezo Voltage Setpoint
Syntax :SOURce:VOLTage:PIEZo?
Vortex Actions Queries the piezo-voltage setpoint.
Response “x.xV”
range: 0.0 to 117.5
Sense Piezo Voltage
Syntax :SENSe:VOLTage:PIEZo
Vortex Actions Returns the piezo-voltage reading.
Response “xxx.xV”
range: 0.0 to 117.5
Vortex6000.book Page 47 Tuesday, February 19, 2002 3:25 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents