PPA5xx/15xx communications manual
2-15
ALARM?
ALARM?
Function:
Read alarm status.
Description:
Reads the status of the measurements
and 2 alarms.
Format:
ALARM?
Arguments:
none
Reply:
single integer
bit 0 data available
bit 1 data error
bit 2 alarm 1
bit 3 alarm 2
Example:
ALARM?
1
Notes:
An alarm is present if bit 0 is high (data is
available) and either alarm 1 or alarm 2
bits are high.