PPA5xx/15xx communications manual
2-17
ALARM2
ALARM2
Function:
Set parameters for alarm2.
Description:
Set alarm2 type and thresholds.
Format:
ALARM2,type,data,high,low
Arguments:
type:
DISABLED
HIGH
LOW
INSIDE
OUTSIDE
data
1-4 for zoom data
high:
high threshold
low:
low threshold
Reply:
None
Example:
ALARM2,LOW,3,0,0.5
Notes:
Both thresholds must be sent even if only
one is used.
There is no LINEAR option for alarm 2.