PPA5xx/15xx communications manual
A-1
Appendix A Configurable parameters
All parameters can be accessed using the CONFIG
command:
CONFIG,number,parameter
number
Function parameter
1
Operating mode,(sets Main Mode)
0=RMS Voltmeter
1=Phase Meter
2=Power Analyser
3=Impedance Analyser
4=Power Integrator
5=Harmonic Analyser
7=Oscilloscope
2
Resolution, (remote options digit resolution)
0=Normal
1=High
2=Binary
4
Autozero manual or auto, (System options)
0=Auto
1=Manual
6
Phase convention, (System options)
0=-180° to +180°
1=0° to -360°
2=0° to +360°
7
Frequency lock on/off, (Acquisition advance options)
0=Off / Normal
1=On / Constant
2=Dynamic
8
Graph, (System options)
0=Dots
1=Lines