PPA5xx/15xx communications manual
2-59
NOOVER
NOOVER
Function:
Disable overranging
Description:
Prevents an overrange error from
blanking out results in manual ranging.
Format:
NOOVER,value
Arguments:
value:
ON
OFF
Reply:
none
Example:
NOOVER,ON
Notes:
This can be useful when testing devices in
a noisy environment. The range can be
set to the correct range for the signal to
be measured even if sporadic noise spikes
would push it up on to the next range.