EasyManua.ls Logo

Newtons4th PPA500 - Phconv

Newtons4th PPA500
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PPA5xx/15xx communications manual
2-65
PHCONV
PHCONV
Function:
Set phase convention and the harmonic
angle.
Description:
Set phase convention and optionally the
harmonic angle.
Format:
PHCONV,convention,angle
Arguments:
convention:
180: -180 to +180
-360: 0 to -360
+360: 0 to +360
Angle:
cosine
sine
Reply:
none
Example:
PHCONV,-360
Notes:
PHCONV,180
PHCONV,180,cosine
0 to -360 degrees is usually used for
power analysis applications.
The Harmonic Angle argument is optional
so does not have to be specified.
However, to update the Harmonic phase
angle argument the phase convention
must be included in the command. See
examples above. The default setting in
the SYS menu is Cosine.

Related product manuals