PPA5xx/15xx communications manual
2-34
DAVER
DAVER
Function:
Set data available enable register
Description:
Sets bits in the data available enable
register to control which status bits set
the data available bits in the status byte.
Format:
DAVER,value
Arguments:
decimal equivalent of data available bits
bit0 set bit 0 of status byte when
new data available
bit1 set bit 0 of status byte when
data available
Reply:
none
Example:
DAVER, 1
set bit 0 in status byte when new data is
available
Notes:
default value is 2:
bit 0 of status byte is set whenever data
is available.