PPA5xx/15xx communications manual
2-85
STATUS?
STATUS?
Function:
Read back channel ranging status.
Description:
Read back condition of selected channel:
range number (1-16)
range text
overflow/underflow status
Format:
or:
STATUS?
STATUS,channel?
STATUS?channel
Arguments:
channel:
CH1
.
.
CH6
Reply:
range number,range text,over/under/ok
1-16
range as per RANGE command
OVER if overflow
LOW if underflow
OK if in range
Example:
STATUS,CH1?
6,300V,OK
STATUS?
OK