PPA5xx/15xx communications manual
2-64
PHASEM?
PHASEM?
Function:
Phase meter query
Description:
Reads phase meter results.
Sets phase meter mode if not already set.
Waits for next unread data if available.
Clears new data available bit read by
DAV?
Format:
PHASEM?
PHASEM,phase?
Arguments:
phase:
PHASE1
PHASE2
PHASE3
PHASES?
Reply:
5 data values separated by commas
freq,mag1,mag2,dB,phase
Example:
PHASEM,CH1
PHASEM,PHASE1?
data returned
Notes:
The phase convention can be set to 0 to
-360, 0 to +360, or +180 to -180 in
the SYSTEM menu or using PHCONV
command.
PHASEM? without specifying the phase
returns the appropriate single phase data.

Questions and Answers: