15
OpenSPR 1-Channel Operation Manual V4.1.01
Figure 2.1 Empty waste bottle installed in the OpenSPR containing 3 outlet tubes.
DI Water Bottle: 250 mL square glass bottle containing a tube fitting
Fill this bottle with 250 mL of filtered (0.2 µm filter) deionized water.
Secure the cap onto the bottle.
Loosen the fitting, and place the 3
rd
inlet tube into the bottle, such that the tube is just
touching the bottom of the bottle.
Gently tighten the fitting on the tube such that it is secured into place.
Place the bottle onto the tray of the OpenSPR next to the Waste Bottle [Figure 2.2].
Buffer Bottle 1 and 2: 250 mL square glass bottles containing a tube fitting
Prepare at least 100-250 mL of each running buffer for your experiment*
Filter (using a 0.2 µm filter) and degass the running buffers prior to use.
Fill the appropriate bottles with the running buffer to a maximum level of 250 mL.
Secure the cap onto the bottle.
Loosen the fitting and place the 1
st
and or 2
nd
inlet tube into appropriate buffer bottle,
such that the tube is just touching the bottom of the bottle.
Gently tighten the fitting on the tube such that it is secured into place.
Place the bottles onto the tray of the OpenSPR [Figure 2.2].