EasyManua.ls Logo

Nicoya OpenSPR - Page 32

Nicoya OpenSPR
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
OpenSPR 1-Channel Operation Manual V4.1.01
Figure 2.33 Prompt after flow cell has been filled with running buffer (sensor successfully detected).
If a signal is successfully detected, perform an inspection at the base of the flow cell for
any accumulation of running buffer and leaking.
Leaking may occur if a proper seal is not made between the sensor and flow cell. If liquid
is present outside of the flow cell, select “Back” to begin the process to re-load the sensor.
If no leaking is present, select “Next” to proceed onto the bubble removal step (Section
2.4.2).
Once a sensor is loaded into the OpenSPR, an absorbance peak should be seen in the graph on
the main screen [Figure 2.34]. The center position of the absorbance peak should be near
approximately 550 nm.
Figure 2.34 Absorbance graph of a sensor installed into the OpenSPR.

Other manuals for Nicoya OpenSPR

Related product manuals