Adjust
so
that
the
phase
plate
image
is
concentric
with
the
annular
diaphragm
image.
Note
that
since
any
displacement
of
the
phase
plate
and
annular
images
will
cause
low
contrast
to
the
phase
contrast
image
of
the
specimen,the
exact
coincidence
of
the
two
images
is
necessary.
System
condenser
turret
unit
Open
the
condenser
aperture
diaphragm
to
the
full.
(The
optical
path
is
blocked
if
the
condenser
aperture
diaphragm
is
narrowed.)
Insert
two
hexagonal
screwdrivers
into
the
annular
diaphragm
centering
screws
of
the
module
in
the
optical
path,
and
rotate
the
screwdrivers
to
center
the
annular
diaphragm.
ELWD
condenser
Unclamp
the
right-hand
and
left-hand
turret
centering
knobs.
Rotate
the
centering
knobs
for
adjustment.
Through
PhL
centering,
other
annular
diaphragms
are
also
centered
automatically.
SLWD
condenser
Bring
the
PhL
annular
diaphragm
into
the
optical
path,
then
rotate
the
centering
screws
for
adjustment.
Maintain
the
field
aperture
diaphragm
fully
opened
during
observation.
(No
centering
adjustment
is
necessary
for
the
Phi
annular
diaphragm
since
it
is
fixed.)