2
PREPARATION
2-8. Face-1/Face-2 Measurement
Face-1 Measurement: Measurements made with the vertical circle positioned to
the left of the telescope eyepiece
Face-2 Measurement: Measurements made with the vertical circle positioned to
the right of the telescope eyepiece
NOTE Be careful not to catch your finger in the opening between the
instrument’s standard and the telescope when rotating the telescope.
Mechanical constant error (except a few errors such as vertical axis error) can be
effectively cancelled out by averaging the values in Face-1 and Face-2.
Face-2 measurement Face-1 measurement
2-14