78
Specifications and Standards
8
8.1
Specifications
Nikon Microscope ECLIPSE 50i
Model ECLIPSE 50i
Optical system Infinity-corrected CF optical system
Objective: CF160
Objectives: Field number 22 (with ergonomic tube/binocular
tube/cytodiagnostic unit), 25 (with trinocular eyepiece tube T/F)
Nosepiece: Sextuple
Focus up/down motion Drive system: Manual coarse/fine motion
(calibration markings for fine motion: 1 µm/marking)
Stroke: 2 mm upward, 28 mm downward
With one-touch refocusing mechanism
Lamp ratings 6V/30W halogen lamp
Lamp type Halogen lamp (PHILIPS 5761)
Average lamp life 100 hours
Input ratings 100-240V AC; 230V (10%, 50/60Hz, 0.9A; 0.5A)
Power cord
100-120V areas:
UL-approved detachable power cord set (3 conductor grounding Type SVT, AWG
18, 3 m long maximum, rated at 125V AC minimum)
220-240V areas:
EU/EN-approved 3-conductor power cord set (3 conductor grounding Type
H05VV-F, 3 m long maximum, rated at 250V AC minimum)
Operating conditions
Temperature: 0 to 40
°
C
Humidity: 85% RH max. (no condensation)
Altitude: 2000 m max.
Degree of pollution: Degree 2
Installation: Category II
Electric shock protection class: Class I
Indoor use only
Transport/storage
conditions
Temperature: -20 to 60
°
C
Humidity: 90% RH max. (no condensation)
External dimensions and
weight (main unit)
External dimensions: 184 (W) x 358 (H) x 383 (D) mm (excluding projections)
Weight: Approx. 9 kg