Chapter 4 Assembly
67
4
(4) Check to confirm that the cytodiagnostic unit
power switch is off.
(5) Firmly insert two cables leading out from the
bottom of the arm into the connectors on the
back of the cytodiagnostic unit.
(6) If the cables are slack and prone to obstruct
operations, secure them with a wire-binding
band provided with the cytodiagnostic unit.
10
Installing Objectives
(when a revolving nosepiece is installed)
Screw objectives into the revolving nosepiece. When installing the objective in this way, make
sure that the magnification of the objective increases when the revolving nosepiece is turned
clockwise (clockwise when viewed from above the eyepiece).
For the 50i: AC adapter cable (thick)
For the 55i: Power cable (thick)
For the 50i: Hand switch cable (thin)
For the 55i: Signal cable (thin)
Bindin
g
band