Chapter 4 Assembly
68
11
Installing Filter Cubes and Light Shield (with the
fluorescence attachment installed)
(1) Remove the cover on the front left of the
fluorescence attachment.
(2) Insert the filter cube.
The filter cubes listed below cannot be
installed directly into the fluorescence
attachment filter bay. The internal spacer
must be removed or reversed, as described
on P. 57.
UV-2A
UV-2B
DAPI
FITC
GFP-L
GFP-B
TRITC
Tx-Red
(3) Insert a nameplate into the position with the
same address as the one indicated on the
filter cube select knob on the right side of the
microscope.
(4) Turn the filter cube select knob and insert a
filter cube into the remaining open bay.
(5) Replace the cover.
Filter cube select knob
Address indication
Name
p
late
For clarit
y
, the e
y
e
p
iece tube is not shown here.
Filter cube
Filter cube ba
y
S
p
acer