Chapter 8 Specifications and Standards
8.1 Specifications
83
8
J-CY Cytodiagnostic Unit for Nikon Microscopes
Model J-CY Cytodiagnostic Unit
Optical system Infinity-corrected CF optical system
Objective: 20X
Variable magnification lens: 2X, 0.5X
Magnification
adjustment system
Quick magnification adjustment by rotary solenoid
Operated by C-HS hand switch
Marker section Quick stamp system (Ink: 7-CY refill ink)
Input voltage 12V DC, 2.5A
Power supply AC adapter
(Power may be supplied directly from the ECLIPSE 55i [through the
cytodiagnostic-unit power supply connector] using the cable set for the 55i &
cytodiagnostic unit when attaching the adapter to ECLIPSE 55i.)
Specified AC adapter Manufacturer: ILAN ELECTRONICS LTD.
Model: F1650K
Rated input voltage: 100-240V AC, 1.2A max.
(when inputting 115V AC), 50/60Hz
Rated output voltage: 12V DC, 3.5A max.
Other: UL-listed product, GS-approved, CE-certified
AC adapter power cord
100-120V areas:
UL-listed detachable power cord set (3 conductor grounding Type SVT, 18 AWG,
3 m long maximum, rated at 125V AC minimum)
220-240V areas:
EU/EN-approved 3-conductor power cord set (3 conductor grounding Type
H05VV-F, 3 m long maximum, rated at 250V AC minimum)
Operating conditions
Temperature: 0 to 40
°
C
Humidity: 85% RH max. (no condensation)
Altitude: 2000 m max.
Degree of pollution: Degree 2
Installation: Category II
Indoor use only
Transport/storage
conditions
Temperature: -20 to 60
°
C
Humidity: 90% RH max. (no condensation)
Weight Approx. 1 kg