10
Product Description
No� Key Name Display Function
TORQUE Key Select a torque limit value� When it reaches the
maximum value, “--” is displayed�
If the torque limit value is to exceed the upper
limit or fall below the lower limit, the alarm
sounds�
If this key is kept pressed, it is fast-forwarded�
The unit is “Ncm”�
PROGRAM Key Select a program number from 1 to 9�
If the PROGRAM key is kept pressed, it is
fast-forwarded�
Adjust the alarm sound volume by keeping
pressing and at the same time� (See “5 4
Alarm sound volume control”�)
Volume high
Volume low
Limited off
Volume OFF: Alarm when rotating in
reverse or when torque limit value
reached
Volume low: Alarm at confirmation or
error
MEMORY key - Preset program settings (rotation speed, torque
limit value, gear ratio, auto reverse mode)� (See
“5 – 3 Programs”�)
- AC IN Lamp - This lamp lights, when the power is supplied from
the AC adaptor, regardless of the ON or OFF
position of the power
- CHARGE Lamp - This lamp lights or flashes while the battery is
being charged or in the refresh mode (see “7 1
Battery refresh”), or in an error condition,
regardless of the ON or OFF position of the power

Table of Contents