19
Operation Procedure
English
7
Press the ON/OFF button (A) to start the
motor handpiece�
(A)
There are 2 operations�
Alternate operation
Press (A) for less than 1 second: Rotation
starts
Press again: Rotation stops
Intermittent operation
Press and hold (A) for 1 second or more:
Rotates while it is pressed
Release: Rotation stop
Operation during motor handpiece rotation
Adjustable: Rotation speed, torque limit value, rotation direction
Non-adjustable: Gear ratio, auto reverse mode, program number
CAUTION
Observe the allowable rotation speed and specified torque that the file manufacturer specifies for
use�
NOTICE
If 10 minutes pass without operation of the ON/OFF button, the power turns off automatically
(Auto power off function)� However, the power does not turn off automatically if the motor
handpiece is in motion, even if it is not used�
5 – 2 Auto reverse function
When the motor handpiece starts and its load reaches approximately half of the preset torque limit value,
the alarm sounds�
The alarm sound changes when the load approaches the torque limit value�
When a further load exceeds the torque limit value, the following 3 modes can be selected:
AUTO REVERSE
The motor handpiece rotates in reverse when the torque limit value is exceeded� When the load is
removed, the motor handpiece returns to forward rotation automatically�
Load within the torque
limit value
Forward Rotation
Further loading
beyond the preset
torque limit value
If the load is kept applied, the file
will rotate in reverse.
When the load is removed,···
Reverse Rotation Forward Rotation