18
5
Operation Procedure
5 – 1 Standard operation
1
Select a program number (1-9)
2
Select the rotation direction�
Forward rotation
Reverse rotation
3
Select the gear ratio of the handpiece to be
used�
4
Set the rotation speed recommended by file
manufacturers�
5
Set the torque limit value recommended by
file manufacturers� (Only for the forward
rotation)
2
3
4
5
6
1
(Setting range)
Gear ratio Rotation speed (min
-1
) Torque limit (N·cm)
20:1 Reduction 100-650 0�8-6�5
16:1 Reduction 120-800 0�6-5�2
10:1 Reduction 200-1,300 0�4-3�2
4:1 Reduction 500-3,200 0�1-1�3
1:1 Drive 2,000-13,000 - -
6
Select the auto reverse mode� For details,
see “5 – 2 Auto reverse function”�
: AUTO REVERSE
: AUTO REVERSE OFF
: AUTO STOP