Post-use Maintenance
6 – 3 Lubrication
Lubricate the contra angle head using the F-Type nozzle and the PANA SPRAY Plus�
1
Shake the PANA SPRAY Plus at 2 or 3 times�
2
Screw the F-Type Spray Nozzle into the PANA
SPRAY Plus by turning it 10 times or so�
3
Attach the F-Type spray nozzle to the rear
end of the contra angle head�
4
Spray until the oil coming out of the tip of the
contra angle head is clear of any debris or
foreign substance�
Spray time: 2 or 3 seconds
Number of spray: 1 time
5
Remove the F-Type spray nozzle from the
contra angle head then remove the F-Type
spray nozzle from the PANA SPRAY Plus�
6
If excess oil comes out of the contra angle
head, leave the contra angle head vertically
for a while�
CAUTION
Be sure to use the PANA SPRAY Plus when lubricating the contra angle head�
Lubricate only contra angle head� Do not lubricate products other than contra angle head�
Shake the PANA SPRAY Plus 2-3 times well before use�
Be sure to securely connect the contra angle head to the F-Type spray nozzle when lubricating the
contra angle head� Otherwise, it may cause improper lubrication to the contra angle head and the
oil may flow backward�
6 – 4 Sterilization
Use a steam sterilizer to sterilize the contra angle head� After each patient treatment, sterilize the contra
angle head as follows�
Sterilizable equipment: contra angle head
1
Insert the contra angle head individually into a sterilization pouch that conforms to ISO 11607-1,
and seal the pouch�
2
Sterilize the contra angle head� Conduct sterilization as follows�
Type Gravity Displacement Pre-Vacuum (Dynamic Air Removal)
Temperature 132°C(0/+4°C) 134°C(0/+4°C)
Full Cycle Time ≥15min 3-18 min
Drying Time ≥30min ≥30min