20
Operation Procedure
AUTO REVERSE OFF
The motor handpiece stops without reverse rotation when the torque limit value is exceeded� In this
instance, the LCD display shows “- - -” and the rotation speed alternately� To rotate it again (forward
rotation), remove the load and press the ON/OFF button twice�
AUTO STOP
The motor handpiece rotates in reverse when the torque limit value is exceeded� When the load is
removed, the motor handpiece stops� If you want it to rotate (forward rotation) again, re-press the ON/
OFF button�
Load within the torque
limit value
Further loading
beyond the preset
torque limit value
If the load is kept applied, the file
will rotate in reverse.
When the load is removed,···
Forward Rotation Reverse Rotation Stop
NOTICE
When reverse rotation is set, this Auto Reverse Function is not activated�
If the battery goes down (when charge indicator mark “
, appears), actual load may not
reach the preset torque limit value (in the case of operation by the battery)� In this case, this auto
reverse function will not be activated� When high torque is required, use the AC adaptor or use
this product in a condition where the battery is not significantly consumed (when charge indicator
mark “
, appears)�
If a load is continually applied to the motor handpiece, it may automatically stop to prevent
overheating (the torque limit value indicates “O�H” on the LCD display�) In this case, leave the
motor handpiece for a while until it cools down� When the display returns to normal, the motor
handpiece can be used�