EasyManua.ls Logo

NSK ENDO-MATE DT2 - 4-6 File Insertion and Removal

NSK ENDO-MATE DT2
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Preparation for Use
English
Disconnecting
When removing the contra angle head, pull it out axially�
CAUTION
Before connecting and disconnecting the contra angle head, turn off the power
Check that the contra angle head is securely assembled to the motor handpiece�
NOTICE
Mount the lubricated contra angle head to the motor handpiece� Calibration cannot be performed
correctly if lubrication is not done� See “4 7 1 Calibration” for calibration�
4 – 6 Inserting and removing the file
Inserting
1
Insert the file until it is correctly secured in
place�
2
Depress the push button and insert the file
into the chuck until the file “notch”
mechanism engages� Release the button�
3
Confirm that the file is secure by gently
pulling and pushing the file without
depressing the push button�
Removing
Depress the push button to open the chuck and remove the file�
CAUTION
Before mounting and removing the file, turn off the power
Always keep the file shank clean� Dirt or debris in the chuck could cause poor file concentricity or
low chuck retention force�
Do not use files with problems listed below as the file may break, seize up or disengage from the
chuck�
Bent, deformed, worn, rusted, broken, deficient file�
File which is cracked on the edge or axis�
Non-ISO (EN ISO) standard, or tampered file�
When replacing files, be sure to wipe any dirt from the file insertion port surface with a dry cloth
before attaching a new file�

Table of Contents