OVP and OCP Accuracy ±(% of output + offset)
OVP 5%+ 0.5V
OCP 5%+ 0.5A
Activation tim
<80ms when maximum output rating
Remote Sensing Capability
Voltage drop 1V/per Lead
Load regulation Add 5mV to drop point +S and +output terminal because of load current
Load voltage
Temperature Coefficient ±(% of output + offset) after a 30-minute warm-up
Voltage 0.01% + 3mV
Current 0.02% + 3mA
Stability ±(% of output + offset) after a 1 hour warm-up
Voltage 0.02% + 1mV
Current 0.1% + 1mA
Transient Response Time
Less than 50usec for output to recover to within 15mV following a change in output
current from full load to half load or vice versa
Voltage Programming Speed
No load Half load
Rising Time ≤ 2ms/15V ≤ 4ms/15V
Falling Time ≤ 5ms/15V ≤ 2.5ms/15V
Command Processing Time(average)
Apply Commands
Setting 20ms
Query 32ms
Output setting Commands
Voltage & Current setting 15ms
Voltage & Current query 32ms
Measurement Commands
Voltage & Current query 32ms
The Other Commands
Setting & Query <35ms
- 83 -