Service Manual RU OHAUS
Page 15 / 53 11/01/2010 Ohaus
CAPACITY 15
(INPUT CAPACITY) INPUT CAPACITY

INC.0.002/0.005
Select incrementif
necessary
select increment

READ THE ZERO

capture zero
ADD WEIGHT 0.000
(input standard weight value
and put on platter)
put standard weight on
platter

DONE capture the weight
SHUT DOWN turn off scale and power
on again
NOTE:
IN THE PROCESS IF GIVE THE MESSAGE “ INIT. CELL...”, WHICH INDICATE THE
FAULT CALIBRATION.