Service Manual RU OHAUS
Page 14 / 53 11/01/2010 Ohaus
4. Calibration
• Open sealing wire; • Screw out sealing screw; • operation:
operation: display: description:
enter the menu
enter calibration
QUIET 0-2
ambience condition,
0=quiet; 1= normal;
2=disquiet
enter to calibrate
PUSH CAL BUTTON
push calibration switch press calibration switch
SELECT GEO 12
(INPUT GEO) input weight GEO (0-31)