Span Calibration (Cont.)
Linearity Calibration
Linearity calibration uses three calibration points; zero, mid-scale and full scale.
Lin Cal must be selected and set to YES in the .S.E.T.U.P. Menu. Before beginning
calibration, make sure the menu Lock Switch is off. Clear the platform.
Start with the balance OFF.
5
1
ON/ZERO
Off
ON/ZERO
Off
PRINT
Unit
1
PRINT
Unit
ON/ZERO
Off
(Place mass value on platform)
ON/ZERO
Off
(Place mass value on platform)
(Remove mass)
ON/ZERO
Off
• • •
1
1
1
1
Y
Y
Y
Y
Y
N