Replacement of Consumables/ Maintenance Manual
4-2
Page
2 Macro (Center) Table Oil Seal: If a vacuum error has occurred
Tools Phillips-head screwdriver (No. 0)
Flat-blade screwdriver (No. 2.6-30)
Special oil seal mounting jig (guide, shaft and collar)
Clean wiper and absolute alcohol
1) Remove the three M2 cross recessed pan head machine screws
using the Phillips-head screwdriver, and remove the macro
(center) table.
2) Insert the flat-blade screwdriver into the oil seal groove. Remove
the oil seal by lifting it from the outside (pulley side).
3) Wind the clean wiper around the flat-blade screwdriver. Moisten
it with commercially available absolute alcohol and clean the
shaft and the pulley side.
Never put absolute alcohol near flames as it is
highly flammable. Take care not to cause a spark
by turning on/off electric appliances or
fluorescent lamps.
4) Attach the special jig for inserting the oil seal.
Set the guide to the stepped section on the top surface of the
pulley.
5) Insert the shaft into the axis hole.
6) Drop a new oil seal into the guide, with the lip facing up.
Flat-blade
screwdriver
Pulley side
Axis
Guide
Shaft
NOTIC