Replacement of Consumables/ Maintenance Manual
4-6
Page
6 L-arm: If a vacuum error has occurred or the L-arm vacuum pad has become damaged
Tools Phillips-head screwdriver (No. 0)
1) Remove the M2 cross recessed pan head machine screws
(four positions for one L-arm vacuum pad).
2) Remove the cover of the L-arm vacuum pad.
3) Replace the L-arm vacuum pad .
4) Attach the cover of the L-arm vacuum pad.
Standard Working Time 15 minutes
7 F-arm: If a vacuum error has occurred or the F-arm vacuum pad has become damaged
Tools Phillips-head screwdriver (No. 0)
1) Remove the M2 cross recessed pan head machine screws
(four positions for one F-arm vacuum pad).
2) Remove the cover of the F-arm vacuum pad.
3) Replace the F-arm vacuum pad .
4) Attach the cover of the F-arm vacuum pad.
Standard Working Time 15 minutes