Replacement of Consumables/ Maintenance Manual
4-4
Page
4 Vacuum Table: If a vacuum error has occurred or the table has become damaged
Tools Phillips-head screwdriver (No. 0)
1) Remove the three M2 cross recessed pan head machine screws
, and replace the vacuum table.
Standard Working Time 10 minutes
5 Oil Seal for Vacuum Table: If a vacuum error has occurred
Tools Phillips-head screwdriver (No. 0)
Flat-blade screwdriver (No. 2.6-30)
Clean wiper and absolute alcohol
1) Remove the three M2 cross recessed pan head machine screws
using the Phillips-head screwdriver, and remove the vacuum
table.
2) Lay the vacuum table upside-down on a flat surface.
3) Insert the flat-blade screwdriver into the oil seal groove. Remove
the oil seal by lifting it from the outside.
Flat-blade