Problem Cause Remedy Page
Picture is too bright.
Exposure correction is set in the +
direction.
Return the exposure correction
value to 0 and set the desired
exposure correction value.
(Online
manual)
The metering area is set to a dark
area outside the region of interest.
Move the metering area to the area
where you want to obtain optimum
exposure.
(Online
manual)
AE lock, which was set when the
exposure time was longer than the
currently required exposure time, is
active.
Cancel AE lock.
(Online
manual)
The input highlight level adjust-
ment is too low.
Reset the current level adjustment
and adjust the optimum level
again.
(Online
manual)
The microscope illumination is too
bright.
Reduce the microscope illumina-
tion intensity or engage an ND filter
to reduce brightness.