151
Chapter 6 Measurement procedures
Modifying specified point
To correct the position of a specified point, select “MODIFY” from the menu in
the stereo measurement screen. The correction confirmation window opens. If
the point to be corrected is in the left image choose “Left Image”. If it is in the
right image choose “Right Image”. Choose “CANCEL” to return to the stereo
measurement screen.
Next, select a point to be correct by tilting [MEAS/ENTER] joystick and press it.
Then move the point by the [MEAS/ENTER] joystick and press it to fix. Press the
[MENU/EXIT] button to cancel correction.
Matching confirmation and distance display
When matching completes, the corresponding point is displayed in input region
in left image. Confirm that the displayed point corresponds to the point specified
in input region in left image. At this time, the distance from the scope distal end
to the specified measurement point is also displayed as “Z= ~ ”. The background
changes yellow when the distance to the measurement point is over 15 mm or
red when it exceeds 30 mm. The distance range up to 15 mm is the
recommended measuring range and 30 mm is the maximum measurable range.
In case auto matching is not smooth
Read “Notes for capturing measurement images” (page 142) and “In case
accuracy cannot be achieved” (page 152). “MODIFY” can also be used to correct
the corresponding point. (Refer to Modifying specified point above.)
Changing the optical adapter (with a retrieved image only)
Select “CHANGE ADAPTER” from the menu to display the stereo optical adapter
list. Correct measurements can be done by changing to the same optical adapter
that was used to record the retrieved image.
Modify window