EasyManua.ls Logo

Omega C9000A SERIES - Example of Programming Functions.19 &.2 (SP2)

Omega C9000A SERIES
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
Ensure that the jumper behind
the lower front bezel is in the
Parameter Setting position
(the jumper should be on the
two right-most pins). Refer to
Section 3.13 if necessary.
To convert the main display
from a temperature read-out
to the Function/Option Mode:
To change from Function .0 to
Function .19:
To display Option:
To change Option value (refer
to Table 3-4); for example, 7:
To change to Function:
3.5.1.2 Example of Programming Functions .19 & .2 (SP2)
STEP ACTION DISPLAY
1
2
3
4
5
6
NOTE
When using SP2 as a tracking second setpoint (deviation from first
setpoint): Function .19, Options 1, 2, 3, or 7 will enable the second
setpoint and Function .2 is to be used to set the differential value
between the first and second setpoint.
Press P
Option 0
Function .0
(flashing)
Press
until you see 0.19
Function (flashing)
Press
Option (flashing)
Press to
increase (or
to decrease
depending on
Option value)
Flashing
Press Function
(flashing)
0.
0
0.
1 9
0.
1 9
7.
1 9
7.
1 9
The following steps apply to Function .2. Function .19 must be set
up before setting Function .2.

Table of Contents

Related product manuals