EasyManuals Logo

Omega C9000A SERIES User Manual

Omega C9000A SERIES
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
75
POWER; Dual 115/230V ±15% 50/60 Hz 5 VA,
Factory set, jumper changeable
OUTPUT MODULES
RELAY ON OUTPUT 1: SPDT relay, rated 5A resistive at
250V continuous use
RELAY ON OUTPUT 2: SPDT relay, rated 3A resistive at
250V continuous use
NOTES
Dual relays are the standard output modules (CN9111A). The relay
ratings shown are conservative. Although the Output 1 relay is
capable of handling a current of 10A and the Output 2 relay a
current of 5A, the life of the relays will be considerably reduced if
the controller is used continuously (24 hours/day) at or near 10A
(Output 1) or 5A (Output 2).
Using the controller in an elevated ambient temperature, or with a
slightly inductive load, also derates the relays. In those situations,
the rating of Output 1 is 5A and Output 2 is 3A.
OPTIONAL OUTPUTS
5VDC SSD: Non-isolated 5 VDC pulses 25 mA for
driving an external DC controlled solid
state relay
1A TRIAC: 1A/264V 100 VA max, SPST
4-20 mA: Isolated, 500 maximum
0-10 VDC: Isolated, 20 mA maximum
ACCURACY & RANGES
CALIBRATION
ACCURACY: ±0.25% of range ±1°C (±0.5°C in
Hi-res) plus linearity tolerance
CONTROL STABILITY: Typically ±0.15% of full scale,
dependent on application
TEMPERATURE
INPUT RANGE: T/C Input-Table 8-1 Millivolt Input-
Table 8-2
LINEARITY
TOLERANCE: T/C Input-Table 8-1 Millivolt Input-
Table 8-2
SECTION 8 SPECIFICATIONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omega C9000A SERIES and is the answer not in the manual?

Omega C9000A SERIES Specifications

General IconGeneral
BrandOmega
ModelC9000A SERIES
CategoryTemperature Controller
LanguageEnglish

Related product manuals