48
Firmware update
The firmware update can be performed only for used devices that are either configurable, in recovery
mode, or require a firmware update. An ongoing verification/sniffing, observation, recording, or just
being armed will prevent the device from being upgraded.
A firmware update is required if the major software version has changed. Minor software versions are
compatible and do not require a firmware update.
In case a firmware update is required, no other actions can be performed.
To manually update the firmware, click Tools > Firmware update.
By default, the firmware update that came with the last installation is used for the actual firmware
update and the corresponding version is displayed. You can use a local firmware image by clearing
the Default image check box and browsing the custom image path.
During the firmware update, you may also need to update the slaves. A slave can be either the
internal slave or an external extension device. The slave is automatically rebooted without the need for
rebooting the device. In case a slave update is required only (represented again through an issue), the
slave update is performed as a single step.
A step-based progress is displayed (finished or error). The following figure shows the firmware update
progress.
Figure 23: Firmware update progress
The whole firmware update may take up to two minutes including reboot and slave updates.
Important:
During the firmware update, the bandwidth is not limited.
Extension devices are automatically updated during the firmware update process if needed.
To avoid an automatic update, unplug the extension devices.