Technical data
65
8.4 Binary inputs
Binary inputs: BINARY INPUT 1...10
Number of binary inputs
10
Trigger criteria
Potential-free or DC-voltage compared to threshold voltage
Sampling frequency
10 kHz
Time resolution
100 µs
Max. measuring time
unlimited
Debounce time
0 … 500 ms (refer to “Debouncing input signals” below)
Deglitch time
0 … 500 ms (refer to “Deglitching input signals” below)
Configuration
Binary inputs can be configured. Refer to DANEO Control Help.
Connection
4 mm/0.16 " banana sockets on the front panel
Insulation
10 galvanic insulated binary inputs. Functional isolation between
with 4 mm creepage between channels.
Reinforced insulation from all SELV interfaces and from power
supply.
Data for potential-free operation
Trigger criteria
Logical 0: R > 80 k
Logical 1: R < 20 k
Input impedance
(125 k ± 20%) || (170 pF ±50 pF)