EasyManua.ls Logo

Omron DEVICENET#05-2003 - Page 171

Omron DEVICENET#05-2003
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
161
GT1-TS04T and GT1-TS04P Temperature Input Units Section 5-3
lected in step 7) on the Temperature Input Unit from the standard voltage/
current generator.
Wait approximately 1 minute for the input to stabilize, then click Fix upper-
limit Value.
The calibrated upper limit will be stored in the EEPROM in the Temperature
Input Unit.
10. Compensate for bias. Turn OFF the standard voltage/current generator,
and enable the cold junction compensator (close the contacts) as shown
below. Check that the cold junction compensator is functioning at 0
°C.
With sensors R, S, and B, the margin of error at close to 0
°C is great, so
set the input type to a K thermocouple to compensate. Wait approximately
40 minutes for the temperature to stabilize, then click Fix Bias Shift Value.
The bias compensation will be stored in the EEPROM in the Temperature
Input Unit.
11. Repeat steps 7 to 10 to adjust other inputs as necessary.
Input type Range (°C) Range (°F) Input voltage
R 0 to 1700 0 to 3000 20 mV
S 0 to 1700 0 to 3000 20 mV
K –200 to 1300 –300 to 2300 50 mV
K 0.0 to 500.0 0.0 to 900.0 20 mV
J 100 to 850 –100 to 1500 50 mV
J 0.0 to 400.0 0.0 to 750.0 20 mV
T –200.0 to 400.0 –300.0 to 700.0 20 mV
L –100 to 850 –100 to 1500 50 mV
L 0.0 to 400.0 0.0 to 750.0 20 mV
B 100 to 1800 300 to 3200 20 mV
Temperature Input Unit
(GT1-TS04T)
Compensating conductor Cold junction compensator
Contacts closed

Table of Contents

Related product manuals