3.2.2.2. CHIP BLOWER
When instrument is selected, by pressing the pedal (A) of the foot control chip-blower is
activated.
TURBINE MODULE
When turbine is selected by taking it out, therefor all others instrument are disable as well as the patient chair
movements are disabled.
The spray is always activated. The water spray flow water can be adjusted by opening/closing the tap located
under the turbine. Wthen turned counter-clockwise, the flow of water to the spray is gradually reduced; when
turned clockwise the flow is gradually increase. Do not adjust the other settings close to the water tap.
The light of the turbine can be switch on/ off by pressing the button on the keyboard.
The air pressure set up for the turbine is calibrated at O.M.S. factory. It should be checked when the
installation of the dental units is done. DO not increase the pressure over the indication of the turbine
manufacturer. This operation must be performed by an authorized O.M.S. technician.
Note:
For further details refer to the manufacturer’s user manual of the turbine.
MICROMOTOR
When micromotor is selected by taking it out, therefor all others instrument are disable as well as the patient
chair movements are disabled.
In normal operation the micromotor rotates in a clockwise direction, the speed can be adjusted by using the
pedal from 900 RPM to 40.000 RPM.
To activated the reverse, press the button on the keyboard.
Note:
Activate the reverse, when micromotor is not working to avoid damages.