OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
22
Spray is enable by pressing on the keyboard. See also MICROMOTOR OPERATION.
The water spray flow water can be adjusted by opening/closing the tap located under the micromotor. When
turned counter-clockwise, the flow of water to the spray is gradually reduced; when turned clockwise the flow
is gradually increase. Do not adjust the other settings close to the water tap.
The light of the turbine can be switch on/ off by pressing the button on the keyboard.
Note:
For further details refer to the manufacturer’s user manual of the turbine.
SCALER
When scaler is selected by taking it out, therefor all others instrument are disable as well as the patient chair
movements are disabled.
Spray is enable by pressing on the keyboard. See also SCALER OPERATION.
The water spray flow water can be adjusted by opening/closing the tap located under the micromotor. When
turned counter-clockwise, the flow of water to the spray is gradually reduced; when turned clockwise the flow
is gradually increase.
EMS scaler:
- Endodontics mode, needs to use a correct tip, and need to adjust the power (between 10 and 30%).
Never increase the power over 30% in Endodontics mode.
- Periodontics mode, needs to use a correct tip, and need to adjust the power (between 10 and 50%).
Never increase the power over 50% in Periodontics mode.
SATELEC SP NEWTRON has to adjust the power according to the tip used.
If the scaler with light has been installed, can be switched on by pressing the button , on keyboard,
Note:
For further details refer to the manufacturer’s user manual of the scaler.