Waveforms:
The Waveform screen
can help the user identify
disturbances to the ow or
if there are any problems
in the signal between the
meters electrodes.
Access this screen through Main Menu --> Diagnostics
Diagnostics:
The user can use the diagnostics menu, coupled with the
alerts menu to help troubleshoot any meter problem. In the
diagnostics screen, a user can access the current waveform,
raw input and output information, network status/ activity, and
signal connections.
Amplitude
of the signal
depends on
the flow rate
No ow without noiseFlow without noice (IDEAL)
Empty pipe (Check installation)
Raw I/O:
The FT-3500 Raw Values screen will show the user the voltage
of electrodes 1 and 2, the raw values for pulse and analog
inputs/ outputs, and the current or resistance from the
temperature sensors (if using the FT-3500 as an energy meter).
High ow with noise (Check grounding)Low ow with noise (Check grounding)