SAMPLE ANALYISIS- WET FLOWER
6. Gather at least 2g of sample from the plant. Note the cannabinoid content may
be significantly different from a leaf sample when compared to a flower. We
recommend gathering flower parts for flowering plants and large healthy leaves
for vegetative plants. Grinding or macerating the sample may improve sample
extraction and is recommended for dried material.
7. Place scale on level surface and turn on.
8. Check calibration of scale by placing 10g calibration weight on scale. Results
should be between 0.997 and 10.003. If not proceed to Section 4 to recalibrate
scale.